- Anesthesia Mode: the module can be paused indefinitely without an alarm.
- Bolus Dose: Allows a bolus infusion to be programmed using either Drug Library or drug calulation feature.
- Callback: A callback for a programmed delay (see "Delay Options" definition) can be scheduled to give an alert Before an infusion is to be initiated, After an infusion is completed, Before and After an infusion, or no alert (None).
- Channel Labels: Provides a hospital-defined list of labels, displayed in Channel Message Display
- Concentration Limits: Limits specified range of concentrations allows for a particular drug in a profile
- Delay Options: Allows for unit to delay infusion by either a preset time or user defined.
- Drug Calculation: Allows entry of drug for a continuous infusion or entry of flow for a continuous infusion
Specifications
-
- Communication Data Port: RS-232 with RJ45 connector
- Dimensions: 6.9"W x 8.8"H x 9"D
- Electrical Classification: Class 1, Internally Power Equipment
- Fluid Ingress Protection: IPX1, Drip Proof
- Mode of Operations: Continuous
- Power Requirements: 100-240V, 50Hz/50Hz, 150VA MAX
- Programmable for continuous, auto or manual intermittent, or volume/time infusion.
- Pump is recommended for intravenous or arterial infusions of blood, lipids, fluids, antibiotic, and any medication requiring precisely-controlled infusion rates
Specifications
- Flow Rate: 0.1-378 ml/hr., depending on syringe size selected
- Bolus Rate: Up to 378.0ml/hr
- KVO: 0.1 - 9.9 ml/hr in increments of 0.1 ml
- Occlusions 35cc syringe:Normal - max. 16.4 PSI with Mono 35 cc syringe; Low - max. 5.96 PSI with Mono 35 cc syringe; Middle - max. 13.8 PSI with Mono 35 cc syringe; High - max. 15.0 PSI with Mono
- Accuracy: ±3%, excluding variations in syringe accuracies
- Power Requirements: AC Adapter or internal battery